Medisafe Sonic Irrigator “Automatic”- floor model cannulated instrument-cleaning system, for cleaning cannulated instruments both inside and out, in a fully validated, clinically tested process.
Our system combines: Automatic fill, pre-washing, auto enzyme injection, ‘pulsed’ sonic-flushing wash,‘pulsed’ sonic-rinsing, auto instrument lubrication, auto drain. Cleaning up to 18 instruments at temperatures up to 130ºF both inside and out, in a fully validated automatic process taking approximately 35 minutes per cycle. Multi variable programmed cycles can be preset for operator convenience.
Special plumbing/drainage required and is responsibility of purchaser (call for installation specifications) standard 110v electrical required.
6.6 gal tank size
43.5” length x 19.5” width x 44.0” high
(add 15” with lid open).
For optimal cleaning performance use recommend 3E-Zyme Triple Enzyme Cleaning Solution (#MEDIS-12041)
Specifications
Sonic Irrigator –Auto #MED1157S
Quality
Manufactured in the UK under Medisafe’s IS09001 Quality
Use
For internal and external cleaning of cannulated / lumened surgical instrumentation
Environmental
For indoor use at altitudes up to 2000m / 6000ft
Operating temperature -5°C / 23°F to 40°C / 104°F
Humidity -80% RH max for temperatures up to 31°C / 88°F,
decreasing linearly to 50% RH at 40°C / 104°F
Designed for Pollution Degree 2, i.e. not totally sealed.
General Description
Medical Device Class I
Operating Process
Ultrasonic activity plus CPE
Applied Standards
BS2745: part 1: 1993
BS2745: part 3: 1993
EN60101-1:2001 (2nd Edition)
EN61010-2-045 (2001)
EN61326: 1997
A1: 1998
EN980
Wash Cycle Parameters
Min. Temp. During Cycle: determined by temp. of water supply
Max. Temp. During Cycle: 60° C / 140° F
Vol. of Dispensed Chemicals
Wash chemical: Detergent 160ml / 5.4oz (approx)
Final rinse chemical: 160ml / 5.4oz (approx)
Electricity Supply
110v AC 50 / 60Hz
Power Consumption: 0.8KVA max.
Termination: 3-pin type earthed plug (fused @ 13A if UK)
Lead Length: 1.83M (6 feet)
Printer
Supply Voltage: single phase 110v 0.1 amps (nominal) AC
Required Utilities
Water Supply
Cold water supply terminating into a Male ¾” NPT National Pipe Thread hose fitting
Hot water supply terminating into a Male ¾” NPT National Pipe Thread hose fitting.
Optimum pressure: 1.5 BAR (22 PSI)
Acceptable Pressure Range:
1.5 – 2.5 BAR (21 – 35 PSI) hot & cold
Acceptable Temp. Range:
Hot: 55°C / 141°F - 75°C / 167°F
Cold 5°C / 41°F– 25°C / 77°F
Hardness CaCo3:
not more than 115 ppm (HTM2030)
Drain
Discharge Orifice Type: 18mm outside dia.
unthreaded hose tail
Position of drain: at least 150mm below rear drain outlet
Max. Flow to Drain: 20 liters / 5.3 gal min
Max. Temp. of Effluent: 60° C / 140° F.
Ventilation
Max. Heat Output of Load: 60° C / 140° F
Weight & Dimensions
Operational weight: 60.5kgs / 133.3lbs
Dimensions: Wash Chamber Volume: 27,000 cm3
Outer Cabinet Dimensions: 43.5” (w) 18.5” (front to back)
add approx 6” for plumbing outlet 44.5” (h)
add approx 15” when lid fully opened
|