The Medisafe Sonic Irrigator (SI) PCF automated cleaning system is validated for cleaning and thermal disinfection of da Vinci EndoWrist® instruments.
Medisafe Sonic Irrigator “PCF”- is the most versitale instrument cleaning system available today. Cleaning and disinfection of complex cannulated instruments both inside and out, in a fully validated, clinically tested process. (see disinfection model option below)
Our system combines; automatic fill, auto pre-washing, auto enzyme injection, auto ‘pulsed’ sonic-flushing wash, auto 'pulsed’ sonic-rinsing, auto drain. Cleaning up to 20 instruments at temperatures up to 195ºF both inside and out, in approximately 35 minutes per cycle .... OR.... use special 'sonic only mode' for walk up convenient use as a standard ultrasonic system.
For a complete list of EndoWrist instruments validated with the Medisafe SI PCF system, click here. For a copy of the IFU supplement that ships with validated EndoWrist instruments, click here.
The PCF system has unequalled convenience and versatility ! NEW Dock & Lock basket system, convenient spray arm (optional), oversized basin, ‘sonic only’ mode for regular instrument cleaning, interlocking cabinet design allows you to combine two PCF systems at anytime for a dual tank, max throughput that can grow with your hospitals cleaning demands.
Special plumbing/drainage/electrical required and is responsibility of purchaser (call for installation specifications)
208v/3phase electrical required.
14.7 gal tank size
35.5” length x 26.0” width x 41.0” high (add 19” with lid open).
For optimal cleaning performance always use recommend
3E-Zyme Triple Enzyme Cleaning Solution (#MEDIS-12042)
Specifications
Sonic Irrigator – PCF
#MED11121STD {w/ standard 20 instrument connection basket
#MED11121S {w/ da Vinci EndoWrist® 10 instrument connection basket
#MED11121D {w/ disinfection capability
Quality
Manufactured in the UK under Medisafe’s IS09001 Quality
Use
For internal and external cleaning of cannulated / lumened surgical instrumentation
Environmental
For indoor use at altitudes up to 2000m / 6000ft
Operating temperature -5°C / 23°F to 91°C / 195°F
Humidity -80% RH max for temperatures up to 31°C / 88°F,
decreasing linearly to 50% RH at 40°C / 104°F
Designed for Pollution Degree 2, i.e. not totally sealed.
General Description
Medical Device Class II - for low or intermediate level disinfection
Operating Process
Ultrasonic activity plus high volume flushing
Applied Standards
BS2745: part 1: 1993
BS2745: part 3: 1993
EN60101-1:2001 (2nd Edition)
EN61010-2-045 (2001)
EN61326: 1997
A1: 1998
EN980
Wash Cycle Parameters
Min. Temp. During Cycle: determined by temperature of water supply
Max. Temp. During Cycle: 91° C / 195° F
Vol. of Dispensed Chemicals
Wash chemical: Detergent 414ml / 14.0oz (approx)
Final rinse chemical: 414ml / 14.0oz (approx)
Electricity Supply
208V / 3phase / 60Hz
24 AMPS per phase max
15K watts
Lead Length: 5 feet
Printer:
Supply Voltage: single phase 110v / 0.1 amps (nominal) AC
Required Utilities
Water Supply
Cold water supply terminating into a Male ¾” NPT National Pipe Thread hose fitting
Hot water supply terminating into a Male ¾” NPT National Pipe Thread hose fitting.
Acceptable Pressure Range:
2.0 – 5.0 BAR (29 –72 PSI) hot & cold
Acceptable Incoming Water Temperature Range:
Hot: 104°F - 140°F
Cold 41°F – 95°F
Hardness CaCo3: less than 110 ppm (HTM2030)
Drain
Type: Single 4” diameter floor drain fitted with suitable traps on building side within 20” of equipment
>OR<
Wall drain not more than 14” from floor is acceptable (no more than 48” from equipment) Wall drain must accept 2.5” collar in diameter and be greater than 8 gallons per min.
Max. Flow to Drain: 8 gallons per min
Max. Temp. of Effluent: 60° C / 140° F.
Ventilation
Max. Heat Output of Load: 91° C / 195° F
Weight & Dimensions
Operational weight
267.5kgs / 587.7lbs
Wash Chamber Volume: 14.75 gal
Outer Cabinet Dimensions:
33.9” (w)
22.8” (front to back) add approx 6” for plumbing outlet
43.7” (h) add approx 15” when lid fully opened
|